M-mode (ultrasound) | Radiology Reference Article | Radiopaedia.org (2024)

Last revised by Andrew Murphy on 23 Mar 2023

Edit article

  • Report problem with article
  • View revision history

Citation, DOI, disclosures and article data

Citation:

Carroll D, Murphy A, Bell D, et al. M-mode (ultrasound). Reference article, Radiopaedia.org (Accessed on 08 Apr 2024) https://doi.org/10.53347/rID-66511

Permalink:

https://radiopaedia.org/articles/66511

rID:

66511

Article created:

22 Feb 2019, David Carroll

Disclosures:

At the time the article was created David Carroll had no recorded disclosures.

View David Carroll's current disclosures

Last revised:

23 Mar 2023, Andrew Murphy

Disclosures:

At the time the article was last revised Andrew Murphy had no financial relationships to ineligible companies to disclose.

View Andrew Murphy's current disclosures

Revisions:

9 times, by 7 contributors - see full revision history and disclosures

Sections:

Imaging Technology

Tags:

ultrasound, echocardiography, pocus

Synonyms:

  • M mode
  • Motion mode ultrasonography

Often utilized for its excellent axial and temporal resolution of structures,M-mode(or motion mode) is a form of ultrasonographyin which a single scan line is emitted, received, and displayed graphically. An M-mode recording is conventionally displayed with the abscissa representing time, and the ordinate showing distance from the transducer, the latter derived from the time delay from echo emission to reflection and detection 6.

A single piezoelectric crystal is recorded and displayed graphically of which represents the acoustic impedance or density of the material encountered. These signals are subsequently displayed as dots, the brightness of which is proportional to the amplitude of reflected waves 3.

Physics

An alternating current directed through one of the crystals in the ultrasound transducer footprint results (via the piezoelectric effect) in its structural deformation, which results in the conduction of compression-rarefaction waves through an adjacent conducting medium. If these transmitted pulses encounter an interface between two structures with a sufficiently disparate acoustic impedance, reflection may occur, which may return to the ultrasound transducer and excite the crystal.

As the speed of sound is relatively constant in soft tissues, roughly 1540 meters/second, the distance at which the transmitted pulse encountered the structure in question may be inferred based on the transit time. The distance from the transducer will be graphed on the vertical (y-axis, ordinate) plane of the rendered output, whereas the time elapsed during recording is represented by the horizontal plane (x-axis, abscissa).

The high sampling frequency, with >90% of the cycle time spent in "receive" mode and pulse transmission at a rate >1000/second, M-mode provides not only excellent temporal resolution but a superior axial resolution to B-mode (i.e. the ability to discern non-contiguous structures in a vertical plane).

ADVERTIsem*nT: Supporters see fewer/no ads

Clinical use

M-mode is utilized for its aforementioned superlatives; the temporal resolution is most useful in delineating the path of structures moving at a high velocity, as well as their timing (e.g. in relation to the cardiac cycle). The axial resolution is also vital in echocardiography, as it allows the resolution of delicate cardiac structures (e.g. valve leaflets) which are difficult to resolve, especially with transthoracic echocardiography. Specific examples include;3

  • M-mode in echocardiography
    • mitral valve excursion
      • axial resolution allows for the vertical distance between the maximal early excursion of the anterior leaflet of the mitral valve and the interventricular septum
        • this is known as E-point septal separation (EPSS) and may be used to estimate ejection fraction
      • temporal resolution allows for the timing of the early and late diastolic excursions (the E and A waves, respectively) to be related to the ECG
        • right ventricular diastolic collapse, a specific sign for the presence of tamponade physiology, may be identified and related to the cardiac cycle based on the mitral valve leaflet motion 2
      • presence or absence of systolic anterior motion (SAM) of the mitral valve with anterior leaflet prolapse into the left ventricular outflow tract
    • excursion of the lateral tricuspid annulus during systole
      • commonly referred to as tricuspid annular plane systolic excursion (TAPSE), an index of right ventricular function1
  • M-mode in lung ultrasonography
    • might be used to record the presence or absence of lung sliding if one is unable to store cine-loops
    • differentiation of a fluid collection, such as a pleural effusion, from intraparenchymal lung pathology
      • fluid dynamics dictate fluids are non-compressible, have a fixed volume, but (unlike solids) do not have a fixed shape
      • respiratory maneuvers will change the shape of the pleural cavity surrounding the fluid, resulting in subtle changes which may be documented in m-mode
        • pleural location identified by the quad sign
        • the sinusoid signis only present in a free-flowing fluid

Related articles: Ultrasound

Promoted articles (advertising)

M-mode (ultrasound) | Radiology Reference Article | Radiopaedia.org (2024)
Top Articles
Latest Posts
Article information

Author: Barbera Armstrong

Last Updated:

Views: 6683

Rating: 4.9 / 5 (79 voted)

Reviews: 86% of readers found this page helpful

Author information

Name: Barbera Armstrong

Birthday: 1992-09-12

Address: Suite 993 99852 Daugherty Causeway, Ritchiehaven, VT 49630

Phone: +5026838435397

Job: National Engineer

Hobby: Listening to music, Board games, Photography, Ice skating, LARPing, Kite flying, Rugby

Introduction: My name is Barbera Armstrong, I am a lovely, delightful, cooperative, funny, enchanting, vivacious, tender person who loves writing and wants to share my knowledge and understanding with you.